Preliminary Testing of a Mach-Scaled Active Rotor Blade with a Trailing Edge Servo-Flap
Steven
R. Hall, Massachusetts Institute of Technology 77
Massachusetts Ave Cambridge, MA 02139-4307 USA, Office:
33-213, E-mail: Steven_Hall@mit.edu ,
Telephone:, 617-253-0869, Fax: 617-253-7397
Eric
F. Prechtl Massachusetts Institute of Technology 77
Massachusetts Ave Cambridge, MA 02139-4307 USA, Office:
37-391, E-mail: prechtl@mit.edu ;
Telephone:, 617-253-3267, Fax: 617-258-5940
Keywords:
helicopter, rotor blade, servo-flap, actuator, piezoelectric
The
results of preliminary tests on an active helicopter rotor blade are presented.
The blade, a Mach-scaled model of a CH-47D helicopter blade, has a discrete
piezoelectric actuator embedded within the spar that controls a trailing edge
flap via a pushrod. Ultimately, the blade will be tested on a helicopter rotor
hover stand at MIT. In this paper, we describe the tests performed prior to
hover testing. First, the actuator was tested on the bench to determine its
control authority and frequency response. Second, the actuator was tested on a
shake table to simulate the out-of-plane accelerations that would be
encountered in a full-scale helicopter in forward flight. Third, the actuator
was embedded in the model blade, and its response to low-frequency sinusoidal
actuation was obtained and compared to the bench test results. Finally, the frequency
response of the actuator in the blade was determined using swept sine
excitation. All test results indicate that the actuator should produce the
desired level of control authority in the model-scale rotor.
Helicopters
are subject to high levels of vibration, mostly due to the unsteady
aerodynamics of the main rotor, especially blade-vortex interaction. The
vibration increases maintenance requirements, reduces pilot effectiveness, and
reduces passenger comfort. Feedback control can be used to reduce this
vibration. Early studies on helicopter rotor control considered the use of
actuators that control the blade pitch through the swashplate, using either
Higher Harmonic Control1 (HHC)
or Individual Blade Control2 (IBC)
methodologies.
During
the last decade, there has been considerable interest in the use of on-blade
actuation for helicopter rotor control for vibration and noise, for a number of
reasons. Generally, HHC and IBC techniques require high bandwidth (on the order
of N times the rotor frequency, where N is the number of rotor blades), and low
blade pitch amplitude (about ±2 deg). On the other hand, the swashplate and the
actuators that drive it are designed for maneuvering control, which is at a
much lower frequency, and requires higher amplitudes. Also, the swashplate is
flight-critical, whereas vibration control is not. Thus, there is some
understandable reluctance to use the swashplate system for vibration control.
Also, several studies point to the use of trailing-edge flaps as an effective
means to control vibration.3,4
Spangler
and Hall5 first
suggested the use of active materials, such as piezoelectric ceramics, for
on-blade actuation. Piezoelectric materials are in some ways ideal candidates
for this application, because of their inherently high bandwidth and stiffness.
Spangler and Hall proposed using a bimorph ender element, cantilevered behind
the blade spar towards the trailing edge, that would control a hinged trailing
edge flap. Later, Hall and Prechtl6
improved on that design with a tapered bender, and a flap
connected to the bender and airfoil with flexures. Chopra et al.7-10
have also investigated the use of piezoelectric bender
actuators on Froude-scaled model rotors, and have begun testing of a Mach-scaled
model rotor with these actuators. Fulton and Ormiston11,12
used similar actuators on a Froude-scaled blade to determine
the response of a model rotor to trailing-edge flap control.
Unfortunately,
bender actuators have a number of limitations. Their placement in the rear of
the airfoil imposes a severe weight penalty on the blade, since the bender mass
must be counter-balanced with weight at the leading edge for aeroelastic
stability. More importantly, the energy density of current piezoelectric
materials, using the d31 effect
required for bending actuation, is about an order of magnitude too low for
effective flap-actuation in full scale or Mach-scale blades. The energy density
of piezoelectric materials using the d33 effect
is about sufficient for rotor control. However, the challenge has been to
develop an efficient amplification mechanism to convert the small strains these
materials produce (1000-2000 mstrain
peak-to-peak) to usable motion. A number of researchers have developed
actuators that amplify the motion of piezoelectric or magnetostrictive stacks.13-17
In
this paper, we report on ongoing research at MIT to develop such an actuator,
known as the X-Frame Actuator, for use in full-scale or Mach-scale model
rotors. Our research results to date are for model rotors. As noted by
Friedmann,18 Mach
scaling is appropriate for tests of model rotors using active materials for
actuation. Indeed, the important nondimensional parameter that relates to the
use of active materials is the ratio of dynamic pressure to material modulus.
The
design considerations and in-depth description of the X-Frame actuator were presented
in two previous papers by the authors.19,20
The basic actuator concept is shown in Figure 1. The X-Frame
actuator consists of two active material stacks and two criss-crossed frames. A
rotational degree of freedom serves as the interface between the two frames at
one end, called the pivot end. As the stacks extend, the frames rotate with
respect to each other about the pivot end of the actuator. The shallow angle
between the stacks and the frames geometrically amplifies the stroke at the
opposite end, called the output end. Restraining one of the frames at the
output end of the actuator causes the opposite frame to undergo the entire
motion.
Refinements
were made to the basic actuator design to facilitate its incorporation into the
Mach scaled rotor blades. These refinements were described in depth in a
previous paper.20 A brief
overview of the design features is provided here. The rotor blade actuation
system is shown in Figure 2.
The
stacks used in the actuator can be composed of any type of active material. The
stacks used in this research were purchased from EDO Corporation. The active
material is designated by EDO as EC-98, which is a lead magnesium niobate-lead
titanate (PMN-PT) composition. Large radius spherical end caps were used on
each stack to allow relative stack/frame rotational motions. Small alignment
pins were also included to maintain stack alignment despite large transverse
accelerations.
The
rotational degree of freedom between the frames is realized through a flexural
connection at the pivot end of the actuator. At the output end, the outer frame
includes guide arms between which the inner frame can slide. These guides both
keep the two frames aligned with each other and help react transverse forces on
the inner frame.
Two
restraints are used to mount the actuator within the rotor blade spar. The
outboard end of the actuator is bolted to the outboard spar restraint,
constraining that end in the flapwise, chordwise and spanwise directions. An
additional flexure, oriented in the spanwise direction, is used between this
bolted mounting point and the actuator frames to equilibrate the centrifugal
loads on the two stacks. The inboard side of the actuator is restrained in the
chordwise and flapwise directions by a sliding interface between the outer
frame and the inboard spar restraint. The actuator is free to move in the
spanwise direction at the inboard side to allow for stack expansion and
contraction.
The
actuator forces and deflections are transferred to the servo-flap by a control
rod. Threads at either end of the control rod are used to connect it to the
inner frame and a clevis at the trailing edge. The clevis/control rod
attachment is accomplished using an 0-80 threaded engagement. This engagement
acts as the adjustment/trimming mechanism for the servo-flap. A pinned
connection is used to connect the clevis to the flap horn, which is bonded to
the servo-flap during its initial cure. A spherical protrusion on the pin is
used to allow for small relative flapwise and chordwise rotations between the
clevis and the horn.
A
graphite reaction rib is laid up in the composite rotor blade, oriented in the
chordwise direction. This rib serves to react actuation forces, as the blade
structure is not normally designed to be stiff in the aft fairing.
A
pre-stress wire runs through the entire length of the servo-flap. It interfaces
with the outboard side of the flap through two keys and with a pre-stress wire
flange at the inboard side. This wire has three functions; the flap rotational
shaft, the flap thrust bearing, and the pre-stress element for the actuator. A
detailed description of how this wire performs these three functions is
presented in the previous paper.20
A
slotted flap was used for the servo-flap design. The airfoil/flap contour was
designed using MSES, a multi-element, viscous, compressible airfoil analysis
code. MSES is a derivative of the single element code, ISES.21-23
The code was also used to compare the redesigned section to
the original CH-47 section. While the slotted flap does develop slightly higher
drag on the blade, it requires substantially lower hinge moments for operation
than a plain flap, which would have a lower drag penalty. The aeroelastic
properties of a slotted flap are also beneficial, since it is easier to mass
balance the flap for stability. Finally, because the slotted flap pivots about
an axis very close to its centroid, the inertia is smaller, which results in
greater actuation bandwidth.
In
this section, we present experimental results with the actuator described
above. Three types of experiments were performed. First, the actuator was
tested on a shake table, to simulate the out-of-plane loads that would be
encountered by a blade-mounted actuator due to unsteady aerodynamic loads.
Second, the actuator was tested in the model rotor blade at low frequency, to
compare its performance to bench top experiments. Finally, the actuator was
tested in the blade with a swept-sine excitation, to determine its frequency
response. The experimental setup and results are described below:
3.1 Shake Tests
One
issue of concern is whether a blade-mounted actuator is sensitive to blade
accelerations. Such a sensitivity can take two forms. First, the actuator may have
induced deflections due to acceleration, which would produce an undesirable
(perhaps even destabilizing) coupling between blade motion and flap deflection.
Second, blade accelerations may cause the actuator to have reduced performance.
For example, increased friction due to inertial forces might reduce the output
deflection. In a typical rotor, the highest (unsteady) accelerations by far
occur in the out-of-plane direction. Therefore, the X-Frame actuator was tested
for acceleration sensitivity by simultaneously operating the actuator while
shaking in the out-of-plane direction at the frequencies and amplitudes that
would occur in forward flight, appropriately scaled to the model rotor.
To
perform the tests, an apparatus was constructed that allowed the actuator to be
operated on a shake table, while simulating as much as possible the conditions
that the actuator experiences in the blade. In particular, inboard and outboard
restraints were manufactured identical to those in the blade, except for flanges
necessary to mount them to the apparatus. To simulate the load of the flap, the
pushrod was connected to a steel load flexure with stiffness matching the
expected aerodynamic hinge stiffness of the flap. Note that the inertia of the
flap was not incorporated into the apparatus, which has implications for the
frequency response testing later.
The
actuator was operated under four different acceleration amplitudes (7.1g,
14.5g, 43g, and 69g); four different acceleration frequencies (22.5 Hz, 45 Hz,
67.5 Hz, and 135 Hz); and four different actuation frequencies (3 Hz, 22.5 Hz,
45 Hz, and 67.5 Hz). In addition, the actuator was operated prior to and after
shake testing with no acceleration. In all, 48 test runs were performed.

Figure 3 shows the results from two of the tests, which are roughly the
two extreme cases. The plot on the left shows actuator excitation at 3 Hz, with
no shaking. The plot on the right shows the result with the same excitation,
and acceleration at 69g amplitude and 45 Hz frequency. In the two cases, the
time responses of the actuator are nearly identical. The peak-to-peak amplitude
of motion is the same, as is the amount of hysteresis due to material
properties. The only slight difference visible is that in the case with
shaking, there is a very small ripple visible, apparently at 90 Hz. We believe
that this is due to the side members of the frames bending at 45 Hz, which
produces foreshortening with fundamental frequency at 90 Hz. In any event, the
ripple is small, and is only noticeable in the highest acceleration cases.
The
results in Figure 3 are typical. That is, all 48 cases have similar time
traces, except for a slight frequency-dependent amplitude variation, probably
due to material properties. The results of all test cases are shown in Table 1.
|
Shake Amplitude (g) |
Shake Frequency (Hz) |
Actuation Frequency (Hz) |
|
|||
|
3 |
22.5 |
45 |
67 |
|
||
|
Actuator Deflection (mil) |
||||||
|
0 |
0 |
15.30 |
14.65 |
14.50 |
14.51 |
|
|
7.1 |
22.5 |
15.05 |
14.44 |
14.30 |
14.28 |
|
|
|
45.0 |
15.11 |
14.48 |
14.33 |
14.28 |
|
|
|
67.5 |
15.12 |
14.51 |
14.33 |
14.29 |
|
|
14.5 |
22.5 |
15.05 |
14.44 |
14.27 |
14.22 |
|
|
|
45.0 |
15.13 |
14.55 |
14.30 |
14.29 |
|
|
|
67.5 |
15.16 |
14.56 |
14.34 |
14.31 |
|
|
|
135.5 |
15.18 |
14.55 |
14.37 |
14.39 |
|
|
43 |
45.0 |
15.09 |
14.50 |
14.19 |
14.12 |
|
|
|
67.5 |
15.09 |
14.52 |
14.31 |
14.31 |
|
|
69.0 |
45.0 |
15.08 |
7.01a |
7.09a |
6.95a |
|
|
0 |
0 |
15.10 |
14.48 |
14.33 |
14.28 |
|
Table 1.
Shake Test Data. All cases were performed with stack voltage amplitudes between
390.5V and 396.5V amplitude. The deflection data in this table is linearly
scaled to obtain the deflection at 400V amplitude.
Note: a
For these cases, the leads to one of the two stacks in the
actuator were inadvertently disconnected. As a result, the amplitude of the
response is approximately one-half that of the other cases.
3.2 Quasistatic Response
Typically,
active materials such as piezoelectric ceramics exhibit nonlinear behavior,
such as hysteresis, as well as voltage-dependent nonlinearities (increasing or
decreasing d33).
In addition, there can be other sources of nonlinear behavior, such as binding or
friction in the actuator mechanism, or nonlinear gain. Thus, it is important to
characterize the quasistatic response of the actuator.

Figure 4 shows the quasistatic responses of the actuator on the bench
top driving a light load (to match that of the prestress wire in the blade),
and in the rotor blade. The data in Figure 4a is normalized to yield equivalent
flap deflection, so that it can be compared directly to Figure 4b. In Figure
4a, the hysteresis is moderate, and is primarily due to material effects. In
contrast, friction in the flap hinge (i.e., the prestress wire) and clevis
increase the hysteresis significantly, reducing the free deflection by about 4
deg. Under aerodynamic loads, the effect of friction will be less important,
due to the increased hinge stiffness of the flap. We expect that we will lose
only about 2 deg of motion due to friction at the aerodynamic design point.
The
friction problem is largely a result of the difficulties associated with
building at model scale. In particular, we were unable to locate suitable
bearings for the hinge line, and as a result our hinge has metal-to-metal
contact without a bearing.
Also,
the friction problem could be reduced significantly in a configuration where
the actuator preload is not applied through the flap itself.
3.3 Frequency Response
Finally,
the transfer functions of the actuator and actuator/flap system were
determined, for two cases, as shown in Figure 5.

Figure
5a shows the frequency response for the actuator alone in the shake test apparatus.
The deflection of the actuator was measured using a strain gage
attached to the load flexure, which had been calibrated using a laser
interferometer. The response is very nearly second order, with natural
frequency at about 650 Hz, and lightly damped. Also, there is slightly
increased phase lag as compared to a second-order system, due to material
hysteresis.
Figure
5b shows the response of the actuator in the blade with the flap attached. The
flap deflection was determined by measuring the actuation deflection using a
Hall effect sensor, which was calibrated by comparing the Hall effect sensor
output voltage to flap angle measured using a laser light lever. The actuator
was actuated using a 0-800 V sinusoidal excitation. Because of significant levels
of friction, the response of the flap was more nonlinear than in the shake test
apparatus. Hence, the data is plotted showing deflection at 800 V excitation,
rather than as a transfer function.
Figure
5b exhibits two natural modes within the frequency range tested. The first mode
is at about 140 Hz. The mode is at a lower frequency than in the shake test
apparatus, which did not include the effect of the flap inertia. Also, note
that this mode is more heavily damped, due to the friction effects. Friction
and hysteresis effects also cause significant phase lag. The second mode is at
about 180 Hz. This mode corresponds to the first torsional mode of the rotor
blade for the test condition, i.e., with the root end free.
Note
that the bandwidth of the actuator in the blade is greater than 6/rev, which is
more than adequate for rotor control on the CH-47, which requires up to 4/rev
actuation.
Finally,
the flap was actuated at several frequencies, and videotaped while illuminated
with a strobe light, in order to visualize the flap motion. Interested readers
can view the video at http://web.mit.edu/srhall/www/blade.html.
The
X-Frame actuator has been incorporated into a Mach-scaled model rotor system,
and preliminary (non-rotating) tests demonstrate that the device has the
expected control authority and bandwidth. The actuator has been shown to work
well in the vibration environment expected in the full-scale rotor. Other
tests, not reported on here, demonstrate that the actuator should work in the
centrifugal field expected. Hover testing of the actuator should begin at MIT
in April 1999.
This
research was supported by DARPA under Contract Number F49620-95-2-0097 and
MDA972-98-3-0001, monitored by Bob Crowe, Bill Coblenz, and Ephrahim Garcia.
Additional support was provided by the Army Research Office, under contract
DAAH04-95-0104, monitored by Gary Anderson. The authors would like to acknowledge
Aaron Julin of MIT and Rich Bussom of Boeing Philadelphia for their support of
the shake test and apparatus; Prof. Mark Drela of MIT for aiding with the
aerodynamic analysis of the flap and airfoil section, as well as code for blade
manufacturing; John Rodgers and Mads Schmidt for assistance with blade
manufacture; and Terry Deane at Advanced Machining and Tooling, Inc. for
machining the model scale actuator components.
1. J. Shaw,
N. Albion, E. J. Hanker Jr., and R. S. Teal, “Higher harmonic control: wind
tunnel demonstration of fully effective vibratory hub force suppression,"
in Proceedings of the 41st Annual Forum of the American Helicopter Society,
pp. 1-15, 1985.
2. N. D. Ham,
“Helicopter individual-blade-control research at MIT 1977-1985," Vertica 11(1),
pp. 109-122, 1987.
3. J. C. Garcia, “Active
helicopter rotor control using blade-mounted actuators," Master's thesis,
Massachusetts Institute of Technology, Department of Mechanical Engineering,
Feb. 1994.
4. T. A. Millot and P. P.
Friedmann, “Vibration reduction in helicopter rotors using an actively
controlled partial span trailing edge flap located on the blade," Tech.
Rep. 4611, NASA, June 1994.
5. R. L.
Spangler and S. R. Hall, “Piezoelectric actuators for helicopter rotor
control," in 31st Structures, Structural Dynamics and Materials
Conference, (Long Beach, CA), Apr. 1990.
6. S. R. Hall
and E. F. Prechtl, “Development of a piezoelectric servoflap for helicopter
rotor control," Smart Materials and Structures 5(1), pp.
26-34, 1996.
7. D. K.
Samak and I. Chopra, “Design of high force, high displacement actuators for
helicopter rotors," in SPIE Proceedings of the Smart Structures and
Intelligent Systems Conference, vol. 2190, pp. 86-98, 1994.
8. C. Walz and
I. Chopra, “Design and testing of a helicopter rotor model with smart trailing
edge flaps," in Proceedings of the 35th Structures, Structural Dynamics
and Materials Conference, Adaptive Structures Forum, 1994. AIAA Paper No.
94-1767.
9. O.
Ben-Zeev and I. Chopra, “Advances in the development of an intelligent
helicopter rotor employing smart trailing-edge flaps," Smart Materials
and Structures 5(1), pp. 11-25, 1996.
10. N. A. Koratkar and I. Chopra, “Testing and validation
of a froude scaled helicopter rotor model with piezo-bimorph actuated trailing
edge flaps," in SPIE Smart Structures and Integrated Systems, pp.
183-205, (San Diego, CA), Mar 1997.
11. M. V. Fulton and R. A. Ormiston, “Hover testing
of a small-scale rotor with on-blade elevons," in Proceedings of the
53rd Annual Forum of the American Helicopter Society, April-May 1997.
12. M. V. Fulton and R. A. Ormiston, “Wind tunnel
results for small scale rotor with on-blade elevons," in Proceedings of
the 54th Annual Forum of the American Helicopter Society, May 1998.
13. R. C. Fenn, J. R. Downer, D. A. Bushko, V.
Gondhalekar, and N. D. Ham, “Terfenol-d driven flaps for helicopter vibration
reduction," in SPIE Smart Structures and Intelligent Systems, vol.
1917, pp. 407-418, 1993.
14. C. M. Bothwell, R. Chandra, and I. Chopra,
“Torsional actuation with extension-torsion composite coupling and a
magnetostrictive actuator," AIAA Journal 33(4), pp. 723-729,
1995.
15. F. K. Straub, “A feasibility study of using smart
materials for rotor control," Smart Materials and Structures 5(1),
pp. 1-10, 1996.
16. E. F. Prechtl and S. R. Hall, “Design of a high
efficiency discrete servo-flap actuator for helicopter rotor control," in SPIE
Smart Structures and Integrated Systems, pp. 158-182, (San Diego, CA), Mar
1997.
17. P. JÄanker, F. Hermle, T. Lorkowski, S. Storm,
and M. Wettemann, “Development of high performance piezoelectric actuators for
transport systems," in 6th International Conference on New Actuators,
June 1998.
18. P. P. Friedmann, “Rotary-wing aeroelastic scaling
and its application to adaptive materials based actuation,"in 39th AIAA
Structures, Structural Dynamics, and Materials Conference, 1998. AIAA Paper
98-2098.
19. E. F. Prechtl and S. R. Hall, “Design of a high
efficiency, large stroke, electromechanical actuator," Smart Materials
and Structures 8, pp. 13-30, 1999.
20. E. F. Prechtl and S. R. Hall, “An X-Frame
actuator servo-flap actuation system for rotor control," in SPIE Smart
Structures and Integrated Systems, (San Diego, CA), Mar 1998.
21. M. Giles and M. Drela, “Two-dimensional transonic
aerodynamic design method," AIAA Journal 25, pp. 1199-1206, Sep
1987.
22. M. Drela and M. Giles, “Viscous-inviscid analysis
of transonic and low Reynolds number airfoils," AIAA Journal 25,
pp. 1347-1355, Oct 1987.
23. M. Drela, “Newton solution of coupled viscous/inviscid multielement
airfoil flows." AIAA 21st Fluid Dyn.,Plasma Dyn. and Lasers Conf., Paper
AIAA-90-1470, June 1990.